
Appendix B: Equipment and Instrument Circuits
201039
One might be tempted to “float” the oscilloscope instead of the circuit being tested. This could
be accomplished by removing the ground prong on the scope’s power plug. However, doing this
would create a safety hazard, since the oscilloscope’s case would be at the same potential as
whatever is connected to the scope probe’s ground. It would then be possible for an individual
simultaneously touching the scope and an earth connection to be electrocuted.
The 3-phase 208 V, 60 Hz, non-isolated supply
Figure B.9 shows the equivalent circuit for the 3-phase, 208 V (rms), 60 Hz, non-isolated supply,
which is also available at terminals on your workbench. This supply is grounded at the source so
that the potential of each of the 3 phases, the “A”, “B”, and “C” phase is 120 V with respect to
ground, but the phase-to-phase voltage (A to B, B to C, C to A) is 208 Volts due to the 120° dif-
ference between each of the phases.
Figure B.9 Equivalent circuit for non-isolated 208 V (rms), 60 Hz 3-phase power supply.
Autotransformer
One electrical source that you will often use for experiments will be a variable transformer (auto-
transformer) plugged into the ISOLATED 120 V (rms), 60 Hz, electrical outlet on the workbench.
This combination is represented by the equivalent circuit shown in Figure B.4. The isolation
transformer itself is a transformer with a turns ratio of one. It is mounted behind the lab bench.
The autotransformer, also referred to by the brand name Variac, enables one to select a portion of
the input voltage to use in a particular experiment. The autotransformer has a single winding that
is connected to a voltage source. By adjusting a slider mechanism (called a wiper), one can vary
the number of turns on the load side of the circuit, hence giving the user control over the magni-
tude of the load-side voltage, from 0 V up to 140 V.
Comentarios a estos manuales