Chapter 5 Specifications
• All specifications apply to measurements with a 10X probe To
verify that the oscilloscope meets specifications, the oscilloscope
must first meet the following conditions:
o The oscilloscope must have been operating continuously
for thirty minutes within the specified operating
temperature.
o You must perform the Do Self Cal operation, accessible
through the Utility menu, if the operating temperature
changes by more than 5° C.
o The oscilloscope must be within the factory calibration
interval
• All specifications are guaranteed unless noted “typical.”
Specifications
1 MΩ ± 2% || 16 pf ± 3 pf
400 V (DC+AC PK-PK, 1 MΩ input
impedance, X10), CAT I
5 Vrms; +/-10 V Peak for 1 s pulse, 50