91
- Select OUT ON
8. CHECK that the measured voltages on the DVM at 10Vp-p, 5Vp-p, 3Vp-p, 1Vp-p, 100 mVp-p and
50mVp-p are in the accuracy range:
5.3.3 Offset Accuracy
Specification: ±1% of programmed value ±10mV, into 50 ohm
Procedure:
1. Connect the OUTPUT connector to the input of the DVM using a 50 ohm coaxial cable and the
50 ohm 0.1% termination.
2. Set the DVM to measure DC volts.
3. Set the unit to generate a SINE wave signal with 1 KHz frequency. Set the unit amplitude to
10mVp-p
4. Set the unit offset to +4V.
7 CHECK that the measured voltages on the DVM at offsets of +4Vp-p, +2Vp-p, 0Vp-p, -2Vp-p,
-4Vp-p are in the accuracy range:
5.3.4 Sine Distortion
Specification: - 65 dB to 20 KHz.
Procedure: