Model: 4076, 4079 50 MHz Arbitrary Function Generator USER MANUAL

Pagina 2

4 waveform, 50 Ω source impedance Reference IN-OUT 10 MHz, TTL compatible, input or output, for external unit synchronization 50 Ω output impedance

Pagina 3 - Safety Summary

94 - Select OUT ON 8. CHECK that the measured voltages on the DVM at 10Vp-p, 5Vp-p, 3Vp-p, 1Vp-p, 100 mVp-p and 50mVp-p are in the accuracy rang

Pagina 4

95 7.14 Set the unit to generate a sine waveform with 1KHz frequency and 10Vp-p. 7.15 Set the distortion analyzer to measure distortion in dB and

Pagina 5 - Table of Contents

96 2. Push the MODE key on the unit front panel. 3. Verify that the oscilloscope displays a 1KHz sine wave. Make sure that the Output is ON.

Pagina 6

97 - Select F2 – FROM, press key 1 and ENTER - Select F2 – DATA, press key 0 and ENTER - Select F5 – EXEC - Select F3 – YES - Select F5 – PREV - Se

Pagina 7 - Section 1

98 Performance Tests Results Setting Minimum reading Maximum reading Reading Frequency DDS 1 MHz 999990 Hz 1000010 Hz Frequency ARB 1 MHz 99995

Pagina 8 - Specifications

99 Yes No Operating Modes: Triggered Burst External Trig

Pagina 9

100 22820 Savi Ranch Parkway Yorba Linda, CA 92887 www.bkprecision.com © 2009 B&K Precision Corp. V02251

Pagina 10 - Modulation Combinations

5 Section 2 Installation 2.1 Introduction This section contains installation information, power requirements, initial inspection

Pagina 11 - Section 2

6 2.6 Power Requirements The Model 4076 and 4079 can be operated from any source of 100 V to 265 V AC, frequency from 48 Hz to 66 Hz. The maxim

Pagina 12 - 2.7 Grounding Requirements

7 DB-9 pin Name Note 1 2 3 4 5 6 7 8 9 - TXD RXD - GND - RTS CRS - - Transmit Data Receive Data - Signal ground - Request to Send Clear to send -

Pagina 13 - 2.10 RS-232 Configuration

8 2.11 GPIB Address The instrument is shipped with the address set to decimal 9. The address can be changed from the front panel by using the &quo

Pagina 14 - 2.12 GPIB Connections

9 Section 3 Operating Instructions 3.1 General Description This section describes the displays, controls and connectors of the Model 4076 and 4

Pagina 15 - Section 3

10 6. Unit Setting Keys - Quick keys for setting units for frequency, time, and amplitude 7. Rotary Knob - Used to increment/decrement numeri

Pagina 16 - 3.2 Display Window

11 4. Menu Functions Display - Displays the menu options available. Use F1-F5 keys on front panel to select the options. 5. Menu Para

Pagina 17 - 3.4 Back Panel Controls

12 Figure 3.3 - Back Panel View 1. Summing In - For CH 1. Use this input to apply an external analog signal to be added to the

Pagina 18 - Model 4079

13 position and width can be programmed at any desired Arbitrary locations. (See section 3.6.3 for details) 6. Ref In/Out - Use this connector t

Pagina 19 - 3.6 MENU Keys

2 SERVICE INFORMATION Warranty Service: Please go the support and service section on our website www.bkprecision.com to obtain a RMA #. Return the

Pagina 20

14 o AMPL| OFST  UNITS (Only when AMPL is selected, press to toggle display in Vp-p, Vrms, dBm)  50 OHM | HI-Z o INTCLK | EXTCLK - WAVE o

Pagina 21

15 o PROT  FROM  TO  ALL  ON | OFF  PREV o SHOW WAVE o PREV • PREV  PREV - MODE o CONT o TRIG  MAN (Manual Trigger)  INT (In

Pagina 22 - 3.6.1 PARAMETER Key

16 3.6.1 PARAMETER Key This key selects and displays the waveform frequency, amplitude, offset and external reference and allows changing the par

Pagina 23 - 3.6.2 MODE Key

17 F4:UNITS - Selects the amplitude units: peak-to-peak, RMS or dBm. F5:50 OHM/HI-Z - Selects the amplit

Pagina 24 - 3.6.3 WAVEFORM Key

18 F4:BURST - (Burst) - Triggers output N cycles for each trigger event, where N ranges from 2 to 999,999. F5: PHASE - Selects the start phase

Pagina 25

19 F4: PULSE - Selects the Pulse waveform and then displays the pulse menu. Refer to section 3.6.5 for details. F5: ARB - Selects the arbitrary

Pagina 26 - 3.6.4 Arbitrary EDIT Menu

20 level output signal at the points specified by address and length. Note: The maximum LENGTH allowed to be set for marker is 4000. Belo

Pagina 27

21 amplitude. Therefore, a value of 8191 corresponds to positive peak amplitude, 0 corresponds to the waveform offset, and -8191 corresponds to the

Pagina 28

22 abort executing the predefined waveform; press YES to execute the predefined waveform. On the NOISE function a menu of ADD and NEW is prompted t

Pagina 29 - 3.6.6 UTILITY Key

23 Protect Function. Protects (makes read-only) a section of waveform memory. F1: FROM - Selects the address of the first point to protect.

Pagina 30 - 3.6.7 SWEEP Key

3 Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be observed during all phases of opera

Pagina 31 - 3.6.8 MODULATION Key

24 F3:INTEN - (Intensity) Selects the intensity of the LCD display. Select a value using the rotary input knob. Valid numeric values are from 1 t

Pagina 32 - 3.6.9 SETUPS Key

25 3. Select between triggered (TRIG), burst (BURST) or gated (GATE) mode. Note: If this is done before turning on sweep, sweep ON selection wil

Pagina 33 - 4,000,000

26 F2: DEV - Defines the FM deviation frequency. F3: SHAPE - Defines the modulation shape, between SINE, TRIANGLE or SQUARE. F4: MOD-FREQ - S

Pagina 34 - 3.10 Power-On Settings

27 Buffer 0 is the factory default setup. F3: STORE - Stores the current front-panel setup to the specified storage buffer. Change the buffe

Pagina 35 - 3.12 Displaying Errors

28 Stored Buffer number Data points of different waveforms (Start address + length) 1 A pts. 2 B pts. 3 C pts. 4 D pts. From the above illustrati

Pagina 36 - 3.13.3 Setting the Output

29 OFFSET 0.00 V Zero offset REPETITION 10 ms Internal trigger rate MODE CONT Waveform mode N-BURST 2 Waves per burst START ADRS 1 Start memory add

Pagina 37 - 3.14 Examples

30 Trig rate short Internal trigger rate too short for wave/burst. Empty location Attempt to restore non existent setting. SC

Pagina 38

31 1. Press PARAMETER to display the Parameter menu. 2. Press F3:AMPL to select the Amplitude parameter. 3. Use the rotary knob to set the amplitud

Pagina 39

32 3.14.3 Creating an Arbitrary Waveform To create a complex arbitrary waveform: * Load a predefined sine waveform * Load a scaled sine waveform

Pagina 40

33 wave crosses the origin). F4:EDIT F2:LINE F1:FROM 251 F2:TO 501 F4:EXEC F3:YES Step 4: Add a negative pulse/glitch (data value -4095) at addres

Pagina 41 - 3.14.8 Using Voltage Offset

4 CAUTION: Before connecting the line cord to the AC mains, check the rear panel AC line voltage indicator. Applying a line voltage other than th

Pagina 42

34 3.14.4 Setting the Arbitrary Frequency The arbitrary waveform frequency is a function of the number of data points used to run the waveform (

Pagina 43 - 4.1 Overview

35 3.14.6 Loading a Waveform into Execution Memory To load a waveform into execution memory, specify its starting address and length in the ARB

Pagina 44 - 4.2 Device State

36 You can store up to 49 front-panel setups in a part of non-volatile Flash known as the settings storage memory. When you recall a stored setup,

Pagina 45 - 4.4 Device Address

37 Section 4 Programming 4.1 Overview 4.1.1 GPIB This section provides det

Pagina 46 - 4.10 Command Syntax

38 User can return to local control with the press of any front panel keys, but it is extremely important to note that this should be done ONLY w

Pagina 47

39 SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, E2, C0 4.4 Device Address The GPIB address of the device may be set to any value from 0 to 3

Pagina 48

40 a) The commands to set the amplitude, the offset, and to switch the output on. The output being switched on is included here in order to preven

Pagina 49

41 The Program Message Unit Separator consists of a semicolon (';'), optionally preceded and/or followed by whitespace characters. A whi

Pagina 50 - 4.11 Status Reporting

42 mnemonics, some Character Data mnemonics have short and long forms. Only the short or the long form may be used. ii) Boolean data Boolean

Pagina 51 - 4.11.5 Error Codes

43 parser how many 8-bit bytes are being sent. Indefinite Form The Indefinite Form has the structure - # - 0 – 8-bit byte – LF^EOI So

Pagina 52

5 Table of Contents Safety Summary ... 3 Section 1 ...

Pagina 53

44 SOURCE:FREQUENCY 3KHZ;:OUTPUT:STATE ON Common Commands may be inserted in the Program Message without affecting the instrument-control c

Pagina 54 - 4.12 Common Commands

45 4.11.2 Service Request Enabling Service request enabling allows the user to select which Status Byte summary messages may cause the device to

Pagina 55

46 A table of error numbers and their descriptions is presented here. No error reported 0 No error Command Errors A command er

Pagina 56

47 Execution Errors An execution error indicates that the device could not execute a syntactically correct command, either since the data were o

Pagina 57

48 :STATus:PRESet command will disable these events from being reported. 401 Power on 402 Operation complete The *OPC co

Pagina 58 - 4.13.1 SOURce Subsystem

49 always be the last in a program message. Command Type: Common Query Syntax: *OPT? Resp

Pagina 59

50 This command is intended for use with overlapped commands. No commands in the instrument are overlapped, and so this command has no effect. Typ

Pagina 60 - 4.13.1.3 Offset

51 e) *SRE - Service request enable command This command sets the Service Request Enable Register bits. Arguments

Pagina 61 - 4.13.1.5 Function

52 *RCL 49 Stored setting location 49 stores the last instrument setting before power d

Pagina 62 - 4.13.1.7 AM modulation

53 :SHAPe SINusoid|SQUare|TRIangle :FREQuency <numeric value> :

Pagina 63 - 4.13.1.8 FM modulation

6 4.5 Message Exchange Protocol ...

Pagina 64

54 Units specifying Arbitrary Point Rate or Wavelength, since the MAXimum or MINimum value is calculated at the time the command is parsed.

Pagina 65 - 4.13.1.9 FSK modulation

55 :VOLT:OFFS 2.5V :VOLT:OFFS MAX Query

Pagina 66

56 4.13.1.6 Point Rate :SOURce:PRATe <point rate> This command is not used. Point rate setting is available for arbitrary mode only.

Pagina 67 - 4.13.1.10 Sweep control

57 4.13.1.7.4 AM FREQuency This command sets the AM modulating waveform frequency Arguments Type:

Pagina 68 - 4.13.1.12 PULSe setting

58 This command sets the FM modulation deviation Arguments Type: Numeric. U

Pagina 69

59 4.13.1.8.5 FM SOURce This command selects the FM modulation source as either internal (then the above settings are effective) or external (and

Pagina 70 - 4.13.2 OUTPut Subsystem

60 Type: Numeric. Units: MHz, KHz, Hz (default) Range: The whole frequency range

Pagina 71 - 4.13.3.1 Trigger Mode

61 4.13.1.10 Sweep control The following commands control the sweep functionality: 4.13.1.10.1 Sweep STATe This command activates or deactivate

Pagina 72 - 4.13.3.3 Burst Count

62 Command Type: Setting or Query Setting Syntax: [:SOURce:]SWEEP:STARt<ws>&

Pagina 73 - 4.13.4 Arbitrary Subsystem

63 4.13.1.12.1 PULSe PERiod This command sets the pulse period to the specified value. NOTE: This also sets the pulse frequency. Arguments

Pagina 74 - 4.13.4.2 Address

1 Section 1 Introduction 1.1 Introduction This manual contains information required to operate, program and test the Model 4076 and

Pagina 75 - 4.13.4.4 Line Draw

64 Command Type: Setting or Query Setting Syntax: [:SOURce:] PULse:RISe <ws>

Pagina 76 - 4.13.4.6 Copy

65 Note: For model 4079, nothing changes in the commands above to control channel 1. But for channel 2, change :OUTP to :OUTP2. For example, to

Pagina 77

66 This command is used to set the trigger mode. It is not a standard SCPI command. Arguments Type:

Pagina 78 - 4.13.4.10 Start Address

67 Examples: :TRIG:BURST? :TRIG:BURS? MAX 4.13.3.4 Internal Trigger Ra

Pagina 79 - 4.13.4.13 Marker Length

68 [:ADDRess] <numeric value> :STATe <Boolean> :LENGth <numeric value> :SAVe Note: For model 4079,

Pagina 80 - 4.13.5 Status Subsystem

69 Response: NR1 4.13.4.3 Data :ARBitrary:DATA <data> This command is used to set the values of the wa

Pagina 81 - 4.13.5.3 Error Queue Enable

70 Considerations: 1) The value of the data at the start and end points must first be set by the user, using the :ARB:DATA command. For e

Pagina 82

71 Syntax: :ARBitrary:COPY<ws><start>,<length>,<destination> Example: :ARB:COPY 1,1000,1001

Pagina 83

72 Options: SINusoid SQUare

Pagina 84 - 4.13.6.3 SCPI Version

73 Example: :ARB:STAR 100 Query Syntax: :ARBitrary:STARt?[<ws>MIN

Pagina 85 - 4.13.6.5 Power-on Buffer

2 - Draw a line between any two points - Clear (set to zero) any set of points or all points - Protect any set of points or all points

Pagina 86 - 4.14.4 LLO - Local Lockout

74 specified by appending a numeric value, to the MARKer keyword. Maximum MARKER Length 4000 points Arguments

Pagina 87 - 4.15 SCPI Command Tree

75 :PTRansition <NRf> :NTRansition <NRf> :EVENt? :ENABle <NRf> 4.13.5.1 Status Preset

Pagina 88 - 4.15.5 :ARBitrary Subsystem

76 Up to 6 ranges may be specified using one :ENABle command, representing the 6 ranges of errors/events. The ranges are then separated from each o

Pagina 89 - 4.15.7 :SYSTem Subsystem

77 Range: 0 to 131,072. Non integer arguments are rounded before execution. Command Type: Setting or Query

Pagina 90

78 4.13.6 System Subsystem The SYSTem subsystem collects the functions that are not related to instrument performance. The functions implemented

Pagina 91

79 This query is used to read the SCPI version to which the instrument complies. Command Type: Query only Query

Pagina 92

80 4.14 IEEE 488.1 Interface Messages 4.14.1 GET - Group Execute Trigger The GET is used by the AWG as a trigger when it is in either the TRIG

Pagina 93 - 4.17.1 General

81 4.15 SCPI Command Tree 4.15.1 Root Node Root [:SOURce] :OUTPut :

Pagina 94

82 4.15.3 :OUTPut Subsystem :OUTPut [:STATe] :SUMming ON | OFF ON | OFF 4.15.4

Pagina 95

83 4.15.6 :STATus Subsystem :STATus :OPERation [:EVENt]?

Pagina 96

3 Output Leakage Approximately 10 mA can be present at the output BNC connector when unit is powered on and the output is off Waveform Characterist

Pagina 97 - Section 5

84 ASCII and GPIB Code Chart Hex Oct Dec ASCII Msg Hex Oct Dec ASCII Msg 00 000 0 NUL 20 040 32 SP MLA0 01 001 1 SOH GTL 21 041 33 ! MLA1 02 002

Pagina 98 - 5.3 Performance Tests

85 Hex Oct Dec ASCII Msg Hex Oct Dec ASCII Msg 40 100 64 @ MTA0 60 140 96 ` MSA0,PPE 41 101 65 A MTA1 61 141 97 a MSA1,PPE 42 102 66 B MTA2 62 142

Pagina 99

86 4.16 Block Transfer (GPIB only) Arbitrary waveform data sent in IEEE488.2 arbitrary block format may take two forms: the definite form and the

Pagina 100 - 5.3.4 Sine Distortion

87 4.17 GPIB Communication Protocol 4.17.1 General This appendix describes the effects of interface messages on waveform generator operation an

Pagina 101 - 5.3.7 Operating Modes

88 The SPE message generates output serial poll status bytes when talk-addressed. SPD-Serial Poll Disable (25 with ATN) The SPD messag

Pagina 102

89 commands are executed. All other GPIB commandsִ setting and operationalִ prompt and error since those commands are under front-panel (local) con

Pagina 103 - 5.3.10 GPIB Capability

90 Basic Listener L4 Unlisten if My Talk Address (MTA) is received Service Request SR1 Complete capability Remote-Local RL1 C

Pagina 104 - Performance Tests Results

91 Section 5 Performance Check Procedures 5.1 Introduction This section provides the procedure for checking the electrical performance require

Pagina 105

92 5.3 Performance Tests The following tests verify that the waveform generator operates and meets specifications. Perform the tests after a war

Pagina 106

93 Amplitude setting Minimum reading RMS Maximum reading RMS DVM reading RMS 10Vp-p 3.493V 3.578V 5Vp-p 1.743V 1.792V 3Vp-p 1.043V 1.078V 1Vp-