14
DC to > 20 kHz minimum bandwidth
Counter Characteristics
Range
50 Hz to 25 MHz
Resolution
Auto ranging, up to 8 digits
Accuracy
± 0.02 % ± 2 digits
Sensitivity
100 mV
rms
typical
General
Memory Storage
Store up to 10
instrument settings
Store up to 20
instrument settings
Power
Requirements
100 V 240 V AC ± 10% (90 V 264 VAC)
Max. Power
Consumption
< 25 VA
< 30 VA
Operating
Temperature
0 °C 50 °C
Storage
Temperature
-10 °C 70 °C
Humidity
95% RH, 0 °C 30 °C
Dimensions
213 mm x 88 mm x 210 mm (WxHxD)
Weight
Approx. 2.5 kg
Safety
Standards
EN55011 for radiated and conducted emissions
EN55082
EN61010
*For square wave, resolution is up to 4 digits when frequency is > 20
kHz.
**For sine wave at 5 V
p-p
into 50 .
Note: All specifications apply to the unit after a
temperature stabilization time of 15 minutes over an
ambient temperature range of 23 °C ± 5 °C.
Specifications are subject to change without notice.
To ensure the most current version of this manual,
please download the latest version here:
http://www.bkprecision.com/search/4040B