Bk-precision 830C - Manual Manual de usuario Pagina 91

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 113
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 90
89
TRIGger:Subsystem:
TRIGger:SOURce < BUS | IMMediate >
Description: Select a source of the trigger
signal.
Parameter:
BUS select a bus command source and enter
the trigger (hold) mode. In this mode, data hold
will be activated to hold each measured reading
after trigger (*TRG) is received from the bus.
IMM select the internal trigger source.
Example: TRIG:SOUR BUS
Note: The AVERage function will be
disabled when entering into trigger mode.
TRIGger:SOURce?
Description: Query the type of the trigger
source.
Response:
Return <Literal> format string: BUS, IMM.
*TRG
Description: Sets instrument to data hold and
return measured reading after trigger.