DESCRIPTION The VOLT_DIV command sets the vertical
sensitivity in Volts/div. The VAB bit (bit 2) in
the STB register is set if an out-of-range value is
entered.
The VOLT_DIV query returns the vertical
sensitivity of the specified channel.
COMMAND SYNTAX <channel>: Volt_DIV <v_gain>
<channel>= {C1, C2, C3, C4}
<v_gain>= 2mV to 10V(or 5V in the CFL
series)
Note: The suffix V is optional.
QUERY SYNTAX <channel> : Volt_DIV
RESPONSE FORMAT <channel>: Volt_DIV <v_gain>
EXAMPLE The following command sets the vertical
sensitivity of channel 1 to 50 mV/div:
Command message:
C1: VDIV 50MV
ACQUISITION
VOLT_DIV, VDIV
Command /Query