7
Safety Symbols
Refer to the operating user manual for warning information to
avoid hazard or personal injury and prevent damage to
instrument.
This symbol shows that the switch is a power switch located at
the front panel. Pressing this button turns on the oscilloscope,
and holding it down for a few seconds turns off the
oscilloscope.
Category I overvoltage conditions.
Measurement instruments whose measurement inputs are
not intended to be connected to the mains supply. The
voltages in the environment are typically derived from a
limited-energy transformer or a battery.
Category II overvoltage conditions.
Measurement instruments whose measurement inputs are
meant to be connected to the mains supply at a standard wall
outlet or similar sources.
Category III overvoltage conditions.
Measurement instruments whose measurement inputs are
meant to be connected to the mains installation of a building.
Maximum Input Voltages
Oscilloscope Inputs
Maximum input voltage to CH1 and CH2 BNC direct(1:1) - CATII 300V
RMS, CATI 150V RMS
Maximum input voltage to CH1 and CH2 BNC via 10:1 probe PR250SA
(included with models 2515 and 2516) - CATII 1000V RMS, CATIII 600V
RMS