Figure 8, which prompts for a “short”. Also, “REL
indicator will be blinking on the display. This is to
prompt the user that the meter is waiting for an action.
At this point, short calibration will be done when the
button is pressed once. Within a few seconds,
the meter will return to normal display and meter will
be calibrated based on the test frequency and settings
selected prior to entering the calibration mode.
Figure 8 - Short Calibration
Quick Procedure
Below is an example of steps to follow to do both
open and short calibration.
1. Select the primary function mode for
measurement (i.e. L/C/R/Z).
2. Select the test frequency for measurement.
Vista de pagina 52