series mode capacitance of 1250pF.
Cs( 1 D × D ) × Cp
Cs( 1 0.5 × 0.5 ) × 1000pF
Cs1250pF
E.G.2: With a me asurement frequency of 1K Hz, a series inductance of
1000uH with a d issipation factor of 0 .5 has a series resistance of
3.14 ohms.
Rs2 × 3.14 × f × Ls × D
Rs2 × 3.14 × 1K × 1m × 0.5
Rs3.14
However, a t a ny given mea surement frequency, the d issipation
factor of a component is the same for both parallel equivalent and
series equivalent circuits.
Additionally, the recipro cal of the dissipation fa ctor (1/D) is
equivalent to the quality factor (Q).
THREE: SPECIFICATIONS
3.1
Power Source
Battery Type: 006P 9V battery.
Power Consumption’s: 155mW
NOTE: Specifications and information are subject to
change without notice. Please visit www.bkprecision.com
for
the most current product information.