Bk-precision 5492BGPIB - Datasheet Manual de usuario Pagina 5

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 5
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 4
Period
Specifications (cont.)
5 www.bk precision.com
True RMS Bench Multimeter
M
odel 5492B
v041014
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
ACV Range Period Ran ge Res olution
Full Scale
Reading
Accura cy
(2)
Input Sensitiv ity
(Sin e Wave)
1
00 mV
to
750 V
1
~ 10 µs
(1)
0
.01 ns
9
.99999
0
.005
1
00 mVrms
10 µs ~ 10 ms 0.1 ns 9.99999 0.005 40 mVrms
10 ms ~ 100 ms 0.1 µs 99.9999 0.01 40 mVrms
100 ms ~ 200 ms 1 µs 199.999 0.05 200 mVrms
(1) If testing frequency is greater than 1MHz, it will be displayed but no specified accuracy is guaranteed.
(2) Specified accuracy at input >5 % of range.
Supplemental
Environmental
Operating Environ men t
0 °C to 40 °C, 32 ºF to 104 ºF
Storage Environment
-40 °C to 70 °C -40 ºF to 158 ºF
Humidity
90 %RH
General
Remote Interfaces
RS232, USB (virtual COM) GPIB (model 5492BGPIB only)
Measurement Category
CAT II 300V, CAT I 1000V, pollution degree 2
Math Functions
Rel, Max/Min/Average/StdDev (for stored readings), dB, dBm, Limit Test, % and mX+b
AC Input
110/220 V ± 10%, 50/60 Hz ± 5%
Powe r Consumption
20 VA
Safety & EMC
EN 61010-1:2001, EN61326-1:2006 and
EN61326-2-1:2006
Dimensions (W×H×D)
8.85” x 3.93 “ x 13.97”
(225 mm×100 mm×355 mm)
Ne t Weig ht
5.51 lbs (2.5 kg)
Three-Year Warranty
Included Accessories
Test leads, AC power cord, spare fuse, user manual, USB cable, calibration certificate
and test report
Frequency
R
esolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
ACV
Range
Frequency Range Resolution
Full Scale
Reading
Accura cy
(2)
Input Sensi tivity
(Sine Wave)
1
00 mV
t
o
750 V
5 ~ 10 Hz 10 µHz 9.99999 0.05 200 mVrms
10 ~ 100Hz 100 µHz 99.9999 0.01 40 mVrms
1
00 ~ 100 kHz
1
mHz
9
99.999
0
.005
4
0 mVrms
100 k ~ 1 MHz
(
1)
1 Hz 999.999 0.005 100 mVrms
(1) If testing frequency is greater than 1MHz, it will be displayed but no specified accuracy is guaranteed.
(2) Specified accuracy at input >5 % of range.
Gate Time at 1 Sec.
Input protection: 750 Vrms or 500 Vdc
Gate Time at 1 Sec.
Input protection: 750 Vrms or 500 Vdc
Vista de pagina 4
1 2 3 4 5

Comentarios a estos manuales

Sin comentarios